| Plage2 /frequency | 24 hours3 | 90 days | 1 an | 2 years | temperature coefficient/℃ 4 |
| Tension DC | TCAL±1℃ | TCAL±5℃ | TCAL±5℃ | TCAL±5℃ |
| 100 mV | 0.0030+0.0030 | 0.0040+0.0035 | 0.0050+0.0035 | 0.0065+0.0035 | 0.0005+0.0005 |
| 1 V | 0.0020+0.0006 | 0.0030+0.0007 | 0.0040+0.0007 | 0.0055+0.0007 | 0.0005+0.0001 |
| 10 V | 0.0015+0.0004 | 0.0020+0.0005 | 0.0035+0.0005 | 0.0050+0.0005 | 0.0005+0.0001 |
| 100 V | 0.0020+0.0006 | 0.0035+0.0006 | 0.0045+0.0006 | 0.0060+0.0006 | 0.0005+0.0001 |
| 1000 V | 0.0020+0.0006 | 0.0035+0.0010 | 0.0045+0.0010 | 0.0060+0.0010 | 0.0005+0.0001 |
| Tension AC efficace vraie (True RMS)2,5,6 |
| Gammes de 100 mV, 1 V, 10 V, 100 V et 750 V |
| 5 Hz - 10 Hz | 0.35+0.02 | 0.35+0.03 | 0.35+0.03 | 0.35+0.03 | 0.035+0.003 |
| 10 Hz-20 KHz | 0.04+0.02 | 0.05+0.03 | 0.06+0.03 | 0.07+0.03 | 0.005+0.003 |
| 20 KHz-50 KHz | 0.10+0.04 | 0.11+0.05 | 0.12+0.05 | 0.13+0.05 | 0.011+0.005 |
| 50 KHz-100 KHz | 0.55+0.08 | 0.60+0.08 | 0.60+0.08 | 0.60+0.08 | 0.060+0.008 |
| 100 KHz-300 KHz | 4.00+0.50 | 4.00+0.50 | 4.00+0.50 | 4.00+0.50 | 0.200+0.020 |
| Résistance7 | Testing current |
| 100 Ω | 1 mA | 0.0030+0.0030 | 0.008+0.004 | 0.010+0.004 | 0.012+0.004 | 0.0006+0.0005 |
| 1 kΩ | 1 mA | 0.0020+0.0005 | 0.008+0.001 | 0.010?0.001 | 0.012+0.001 | 0.0006+0.0001 |
| 10 kΩ | 100 μA | 0.0020+0.0005 | 0.008+0.001 | 0.010+0.001 | 0.012+0.001 | 0.0006+0.0001 |
| 100 kΩ | 10 μA | 0.0020+0.0005 | 0.008+0.001 | 0.010+0.001 | 0.012+0.001 | 0.0006+0.0001 |
| 1 MΩ | 5 μA | 0.002+0.001 | 0.008+0.001 | 0.010+0.001 | 0.012+0.001 | 0.0010+0.0002 |
| 10 MΩ | 500 nA | 0.015+0.001 | 0.020+0.001 | 0.040+0.001 | 0.060+0.001 | 0.0030+0.0004 |
| 100 MΩ | 500 nA‖10 MΩ | 0.300+0.010 | 0.800+0.010 | 0.800+0.010 | 0.800+0.010 | 0.1500+0.0002 |
| DCcurrent | Chute de tension de la résistance interne |
| 100 μA | <0.11V | 0.010+0.020 | 0.040+0.025 | 0.050+0.025 | 0.060+0.025 | 0.0020+0.0030 |
| 1 mA | <0.11V | 0.007+0.006 | 0.030+0.006 | 0.050+0.006 | 0.060?0.006 | 0.0020+0.0005 |
| 10 mA | <0.05V | 0.007+0.020 | 0.030+0.020 | 0.050+0.020 | 0.060+0.020 | 0.0020+0.0020 |
| 100 mA | <0.5V | 0.010+0.004 | 0.030+0.005 | 0.050+0.005 | 0.060+0.005 | 0.0020+0.0005 |
| 1 A | <0.7V | 0.050+0.006 | 0.080+0.010 | 0.100+0.010 | 0.120+0.010 | 0.0050+0.0010 |
| 3 A | <2.0V | 0.180+0.020 | 0.200+0.020 | 0.200+0.020 | 0.230+0.020 | 0.0050+0.0020 |
| 10 A8 | <0.5V | 0.050+0.010 | 0.120+0.010 | 0.120+0.010 | 0.150+0.010 | 0.0050+0.0010 |
| Capacité15 |
| 1.0000 nF | 0.50+0.50 | 0.50+0.50 | 0.50+0.50 | 0.50+0.50 | 0.05+0.05 |
| 10.000 nF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| 100.00 nF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| 1.0000 μF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| 10.000 μF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| 100.00 μF | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.40+0.10 | 0.05+0.01 |
| Plage2/frequency | 24hours3 | 90days | 1year | 2years | temperature coefficient/℃4 |
| TCAL ± 1℃ | TCAL ± 5℃ | TCAL ± 5℃ | TCAL ± 5℃ |
| Courant AC efficace vrai (True RMS)2,6,9 | Chute de tension de la résistance interne |
| Gammes de 100 μA, 1 mA, 10 mA et 100 mA | <0.011,<0.11, <0.05,<0.5V |
| 3 Hz - 5 kHz | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.015+0.006 |
| 5-10kHz | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.030+0.006 |
| (typicalvalue) |
| 1Arange | <0.7V |
| 3Hz-5kHz | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.015+0.006 |
| 5 kHz - 10 kHz (valeur typique) | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.10+0.04 | 0.030+0.006 |
| 3 A Range | <2.0V |
| 3 Hz - 5 kHz | 0.23+0.04 | 0.23+0.04 | 0.23+0.04 | 0.23+0.04 | 0.015+0.006 |
| 5 kHz - 10 kHz | 0.23+0.04 | 0.23+0.04 | 0.23+0.04 | 0.23+0.04 | 0.030+0.006 |
| (Typicalvalue) |
| 10 A Range8 | <0.5V |
| 3 Hz - 5 kHz | 0.15+0.04 | 0.15+0.04 | 0.15+0.04 | 0.15+0.04 | 0.015+0.006 |
| 5 kHz - 10 kHz (valeur typique) | 0.15+0.04 | 0.15+0.04 | 0.15+0.04 | 0.15+0.04 | 0.030+0.006 |
| Connectivity |
| 1 kΩ | 0.002+0.030 | 0.008+0.030 | 0.010+0.030 | 0.012+0.030 | 0.0010+0.0020 |
| DiodeTesting10 |
| 5 V | 0.002+0.030 | 0.008+0.030 | 0.010+0.030 | 0.012+0.030 | 0.0010+0.0020 |
| DCrate(Typicalvalue) |
| (Précision d’entrée normalisée) + (précision de référence normalisée) |
| Température11 |
| PT100 (DIN/IEC751) | Theprobe accuracy+0.05℃ |
| 5 kΩ thermistor | Theprobe accuracy+0.1℃ |
| Fréquence : précision ± (% de lecture)12.13 |
| Gammes de 100 mV, 1 V, 10 V, 100 V et 750 V 14 |
| 3 Hz-10 Hz | 0.1 | 0.1 | 0.1 | 0.1 | 0.1 |
| 10 Hz-100 Hz | 0.03 | 0.03 | 0.03 | 0.03 | 0.035 |
| 100 Hz-1 KHz | 0 | 0.008 | 0.01 | 0.01 | 0.015 |
| 1 KHz-300 KHz | 0.002 | 0.006 | 0.01 | 0.01 | 0.015 |
| Onde carrée15 | 0.001 | 0.006 | 0.01 | 0.01 | 0.015 |
| Erreur supplémentaire du temps de porte ± (% de lecture)13 |
| Fréquence | 1second | 0.1second | 0.01second |
| 3 Hz-40 Hz | 0 | 0.2 | 0.2 |
| 40 Hz-100 Hz | 0 | 0.06 | 0.2 |
| 100 Hz-1 KHz | 0 | 0.02 | 0.2 |
| 1 KHz-300 KHz | 0 | 0.004 | 0.03 |
| Onde carrée15 | 0 | 0 | 0 |
1. En DC : les spécifications sont valables après 60 minutes de préchauffage, avec une période d’intégration de 10 ou 100 NPLC et le zéro automatique activé. En AC : les spécifications sont valables après 60 minutes de préchauffage, avec le filtre AC lent et une onde sinusoïdale.
2. Toutes les gammes ont un dépassement de 20 %, sauf 1000 V DC, 750 V AC, 10 A DC, 3 A AC, 10 A AC et le test de diodes.
3. Par rapport à l’étalon de calibration
4. En dehors de la plage TCAL ± 5 ℃, chaque degré (℃) ajoute un coefficient.
5. Les spécifications sont valables avec une entrée sinusoïdale > 5 % de la gamme et > 1 mV RMS. La gamme 750 V AC est limitée à 8 × 10⁷ V·Hz.
Pour des entrées entre 1 % et 5 % de la gamme et une fréquence < 50 kHz, ajoutez une erreur de 0,1 % de la gamme. Entre 50 kHz et 100 kHz, ajoutez 0,13 % de la gamme.
6. Réponse basse fréquence : trois filtres sont disponibles (3 Hz, 20 Hz et 200 Hz). Au-delà de la fréquence du filtre choisi, aucune erreur supplémentaire n’apparaît.
7. Les spécifications sont valables pour les mesures de résistance 4 ou 2 fils (offset remis à zéro). Sans la fonction de zéro mathématique, la mesure 2 fils présente une erreur supplémentaire de 0,2 Ω.
8. La gamme 10 A n’est disponible que sur le connecteur avant. Chaque ampère ajoute 2 mA de courant de base, ou courant d’entrée > 5 A RMS.
9. Les spécifications sont valables avec une entrée sinusoïdale > 1,5 % de la gamme et > 10 μA AC. 10. Les spécifications portent sur la tension mesurée à l’entrée ; le courant de test typique est de 1 mA et ses variations modifient la chute de tension de la jonction de la diode.
11. La sonde choisie limite la plage et l’erreur de mesure réelles. La précision de la sonde inclut toutes les erreurs de mesure et de conversion de température ITS-90. Le R0 de la PT100 peut être réglé sur 100 Ω ± 5 Ω pour éliminer l’erreur propre de la sonde.
12. Sauf indication contraire, les spécifications sont valables après 60 minutes de préchauffage, avec une entrée sinusoïdale et un temps de porte de 1 s (7 chiffres).
13. Valable pour des entrées sinusoïdales et carrées ≥ 100 mV. Pour des entrées de 10 mV à 100 mV, multipliez le % d’erreur de lecture par 10.
14. La plage va de 10 % à 120 % en dessous de 750 V AC.
15. L’entrée en onde carrée est spécifiée entre 10 Hz et 300 kHz.